点击这里给我发消息
  点击这里给我发消息
点击这里给我发消息
点击这里给我发消息
产品标题:光致变色透镜测量系统
 首 页 | 公司简介 | 新闻中心 | 产品频道 | 产品目录 | 代理品牌 | 诚聘英才 | 留言反馈 | 联系我们    
产品列表
光电测量产品
激光系列产品
生化仪器产品
元件耗材
光通信仪器
脉冲发生器
石墨烯系列产品
光催化光源/光化学
光学影像
探测器
太赫兹仪器

光电测量产品 >> 光谱系统 >> 光致变色透镜测量系统  

This system is designed to perform transmission tests on photochromic lenses as specified in EN1836:1997.

Lenses up to 75mm diameter can be tested over a wavelength range of 280nm to 780nm in the faded and darkened state.

A conditioning beam as specified in the standard is included. The conditioning beam can be set to 50klux, 15klux or switched off, all by the control software.

A thermoelectric heat pump, controlled by the PC, is used to set the temperature required for each measurement and maintain the sample temperature to within ± 0.2°C at the specified levels.

The system includes enhanced BenWin+ software which, amongst other things, will return:-

  • t0 Luminous transmittance in the faded state
  • t1 Luminous transmittance in the darkened state
  • t0 / t1 Faded state / darkened state ratio
  • tW Luminous transmittance at low temperature
  • tS Luminous transmittance at high temperature
  • ta Luminous transmittance at reduced radiation
  • tF (280< l <315) Max &#118alue of spectral transmittance between 250nm and 315nm
  • tF (315< l <350) Max &#118alue of spectral transmittance between 315m, and 350nm
  • tSUVA Maximum &#118alue of solar UVA transmittance between 315nm and 380nm
  • tsign (red, green, blue, yellow) and Q for each
  • tsb Solar blue transmittance

The rate of darkening and fading is also monitored at a single wavelength or over the whole spectrum if required.

The optical part of the system is mounted on an optical bread-board approximately 75cm x 45cm. The electronics (excluding the PC) is contained in a floor standing 19" rack.

Main System Requirements

The main points to be accommodated by such a system are as follows:-

1) The level of radiation in the analysing beam should be high enough to allow fast and accurate measurement of transmission, but not so high as to cause any measurable darkening of the lens during the measurement of the faded state.

2) The optical arrangement must allow for some optical power in the sample whether caused by non-parallelism in the faces of a plane sample or by curvature in a full lens.

3) The temperature of the sample needs to be controlled precisely, even when it is being irradiated at 50klux.

4) The ability of the detector to “see” light scattered from the conditioning beam should be minimised and in any case such scatter must not contribute to the measured &#118alues of transmission.

5) The spectral shape of the conditioning beam should correspond as closely as possible to the specification given in the standard (Moon, etc., section 6.1.3.1, EN1836:1997. Curve attached). It should be uniform over the region of the sample used by the analysing beam and it must be possible to vary the level without changing the spectral shape.

6) The light source, monochromator and detector must allow measurements over the spectral range 280nm to 780nm. The scattered light at 280nm must be sufficiently low to allow measurement of low transmission in this region.

7) The system should be fully automated requiring only that the operator inserts the sample and keys the computer to execute a predetermined sequence of measurements. Complete automation facilitates the addition of automated sample changing if required.

This Bentham system achieves all of the above. Contact us for a complete technical description of this impressive system.

点击数:5277  录入时间:2007/7/28 【打印此页】 【返回
 
 
 

长春博盛量子科技有限公司 版权所有   Copyright(c) 2007 by www.china-quantum.com
电话:+86-(0)431-85916189/85916095 传真:+86-(0)431-85956117

邮箱: info@china-quantum.com   地址:长春市高新开发区繁荣路5099号